Avery Dennison
MED 5741
Single-Coated Nonwoven
MED 5741 is a single-coated, white polyester nonwoven with an acrylic adhesive designed for long-term wear. This product is soft, conformable, and is highly breathable.
Typical Application
Designed for multi-week wear time applications such as glucose monitors, insulin pumps, activity trackers, electrocardiogram monitors and other mobile health and wellness products.
Features and Benefits
- Highly breathable
- Conformable and flexible
- Demonstrated multi-week wear
- Biocompatible per ISO 10993
- Die-cuttable and easy to handle
- Sterilization compatibility with Ethylene Oxide (EO) and Gamma Irradiation
Physical Properties (Not intended as a specification.)
|
Test Method**
|
Typical Values*
|
---|---|---|
Peel Adhesion on Polyethylene (PE)
|
TDS-01
|
0.9 lbf/in 3.9 N/25 mm
|
Peel Adhesion on Stainless Steel (SS)
|
TDS-04
|
2.0 lbf/in 8.8 N/25 mm
|
Quick Stick
|
TDS-05
|
0.7 oz/in 3.1 N/25 mm
|
90 Degree Liner Release
|
TDS-06
|
200 gf/2 in 3.5 oz/in
|
Static Shear
|
TDS-12
|
260 minutes
|
MVTR
|
TDS-16
|
1000 g/m2 per 24 hr
|
Product Construction
|
Typical Values*
|
Description
|
---|---|---|
Carrier Thickness
|
13 mils 330 µm
|
White polyester (PET) nonwoven
|
Adhesive Thickness
|
2 mils 51 µm
|
Acrylic adhesive designed for medical applications
|
Release Liner Thickness
|
3 mils 76 µm
|
White super-calendered kraft
|
MVTR, moisture vapor transmission rate.
*Refer to product specifications for material acceptance limits.
**Test method information available upon request.
STORAGE AND SHELF LIFE
Two years when stored at 23°C (73°F), 50% relative humidity, out of direct sunlight, in original packaging.