Avery Dennison
MED 5740

Single-Coated Nonwoven

MED 5740 is a single-coated, white polyurethane nonwoven with an acrylic adhesive designed for long-term wear. This product is soft, conformable and breathable.


Typical Application

Designed for multi-week wear time applications such as glucose monitors, insulin pumps, activity trackers, electrocardiogram monitors and other mobile health and wellness products. 

documents

Technical Data sheet

Features and Benefits
  • Breathable
  • Conformable and flexible
  • Demonstrated multi-week wear
  • Biocompatible per ISO 10993
  • Die-cuttable and easy to handle
  • Sterilization compatibility with Ethylene Oxide (EO) and Gamma Irradiation
Physical Properties (Not intended as a specification.)
Test Method**
Typical Values*
Peel Adhesion on Polyethylene (PE)
TDS-01
0.7 lbf/in 3.1 N/25 mm
Peel Adhesion on Stainless Steel (SS)
TDS-04
2.1 lbf/in 9.2 N/25 mm
Quick Stick
TDS-05
1.6 lbf/in 7.0 N/25 mm
90 Degree Liner Release
TDS-06
190 gf/2 in 3.4 oz/in
Static Shear
TDS-12
600 minutes
MVTR
TDS-16
400 [g/m2/24 hours]
Product Construction
Typical Values*
Description
Carrier Thickness
10 mils 260 µm
White thermoplastic polyurethane (TPU) nonwoven
Adhesive Thickness
2 mils 51 µm
Acrylic adhesive designed for medical applications
Release Liner Thickness
3 mils 76 µm
White super-calendered kraft

MVTR, moisture vapor transmission rate.
*Refer to product specifications for material acceptance limits. **Test method information available upon request.

STORAGE AND SHELF LIFE

Two years when stored at 23°C (73°F), 50% relative humidity, out of direct sunlight, in original packaging.